Allow the test, specimen, buffer and/or controls to reach room temperature (15-30°C) prior to testing.
1. Bring the pouch to room temperature before opening it. Remove the test cassette from the sealed pouch and use it as soon as possible.
2. Place the cassette on a clean and level surface.
For Serum or Plasma specimen:
Hold the dropper vertically and transfer 3 drops of serum or plasma (approximately 75mL) to the specimen well of test Cassette and start the timer. See illustration below.
For Venipuncture Whole Blood specimen:
Hold the dropper vertically and transfer 3 drops of whole blood (approximately 75mL) to the specimen well, then add 1 drop of buffer (approximately 40mL), and start the timer. See illustration below.
For Fingerstick Whole Blood specimen:
To use a capillary tube: Fill the capillary tube and transfer approximately 75mL of fingerstick whole blood specimen to the specimen well of test cassette, then add 1 drop of buffer (approximately 40mL) and start the timer. See illustration below.
3. Wait for the colored line(s) to appear. Read results at 10 minutes. Do not interpret the result after 20 minutes.