Directions For Use
1. Bring the pouch to room temperature before opening it. Remove the device from the foil pouch and label it with date and patient’s identifications.
2. Insert the sponge end of the saliva swab into mouth. Actively swab the inside of the mouth and tongue to collect oral fluid.
3. Remove the saliva swab from mouth when the sponge is filled with saliva and becomes soft, or the indicator turns blue.
4. Insert the saliva swab into the test cassette holder and push it down. The bump at the end of the saliva swab must be clicked into the hole of the test cassette holder.
5. As the test begins to work, the purple color move across the result window in the center of the test device. Start the timer.
6. Read the results of drug tests at 5 minutes. Do not read the test results after 10 minutes.

Test Item and Cut-off(ng/mL) as follows
| No. | Drug | Abbreviation | Cut-off (ng/mL) |
| 1 | Amphetamine | AMP | 40/50 |
| 2 | Benzodiazepine | BZO | 10/50 |
| 3 | Cocaine | COC | 20/30/50 |
| 4 | Ecstasy | MDMA | 40/50 |
| 5 | Ketamine | KET | 50/100 |
| 6 | Marijuana | THC(metabolite/parent) | 15/30/50 |
| 7 | Methamphetamine | MET | 40/50 |
| 8 | Opiate | OPI | 25/40/50 |
| 9 | Methadone | MTD | 30/50 |
| 10 | Barbiturate | BAR | 50 |
| 11 | Phencyclidine | PCP | 10 |
| 12 | Tramadol | TRA | 20 |
| 13 | Buprenorphine | BUP | 5 |
| 14 | Oxycodone | OXY | 20/40 |
| 15 | Fentanyl | FTY | 20/50 |
| 16 | Alcohol | ALC | 0.02% |